Posting for a customer:
BQ27510-G2 question and datasheet correction. Yes, I need to update it to a G3, but extracting data from it first.
http://www.ti.com/lit/ds/symlink/bq27510-g2.pdf
1. What is the floating point format used for CC Gain and CC Delta? It doesn’t appear to be IEEE 754 or the byte order is mangled.
I am getting very large values that are out of range. Here is debug showing 32 byte read of calib data and subsequent decoding.
Double checked CC Gain = 0x7f71205c hex to floating point conversion and it shows same large value.
http://www.h-schmidt.net/FloatConverter/IEEE754.html
2. Datasheet page 17 has incorrect offsets values for Configuration Charge Termination. Here are correct offsets: